METHODS AND HARDWARE OF DOSE OUTPUT VERIFICATION FOR DYNAMIC RADIOTHERAPY

Abstract

The design of special verification phantom for dynamic radiotherapy checking is described. This phantom permits to insert the dose distribution cross-calibration before every days patients irradiation on Linac with RapidArc. Cross-calibration factor is defined by approximation of large number correction factors measured in phantom at different angles of gantry rotation and middle quantity calculation. The long range stability of all correction factors have been evaluated during checking of setting treatment plans and did not exceed 1 %. The fast checking of patient’s dynamic irradiation with developed method gives possibility texactly realize of planning dose distribution and to increase the capacity of accelerators.

About the Authors

Y. V. Tsitovich
РНПЦ онкологии и медицинской радиологии им. Н.Н. Александрова, г. Минск
Belarus


A. I. Hmyrak
РНПЦ онкологии и медицинской радиологии им. Н.Н. Александрова, г. Минск
Belarus


A. I. Tarutin
РНПЦ онкологии и медицинской радиологии им. Н.Н. Александрова, г. Минск
Belarus


M. G. Kiselev
Белорусский национальный технический университет
Belarus


References

1. IMRT commissioning: Multiple institution planning and Dosimetry comparisons / Report AAPM Task Group 119. – 2009.

2. Jørgensen Mai-Britt K. Machine-specific QC in Aarhus using EPID / K. Mai-Britt Jørgensen, L. Hoffmann, L.P. Muren [et al.] / Proceed of RA workshop // Denmark 2012.

3. Hoffmann, L. Patient specific QC in Aarhus / L. Hoffmann, U.V. Elstrøm, K. Mai-Britt Jørgensen / Proceed of RA workshop // Denmark 2012.

4. Fredh, A. RapidArc QA at Rigshospitalet / A. Fredh, M. Hörling // Proceed of RA workshop // Denmark 2012.

5. Nicolini, G. GLAaS: An absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications / G. Nicolini, A. Fogliata, E. Vanetti [et al.] // Med. Phys. – V. 33. – 2006. – P. 2839.

6. Nicolini, G. The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy / G. Nicolini, E. Vanetti, A. Clivio [et al.] // Radiation Oncology. – V. 3. – 2008. – P. 24–32.

7. Essers, M. Commissioning of a commercially available system for intensitymodulated radiotherapy dose delivery with dynamic multileaf collimation / M. Essers, M. De Langen, M.L.P. Dirkx [et al.] // Radiotherapy and Oncology. – V. 60. – 2001. – P. 215–224.

8. Absorbed dose determination in external beam radiotherapy / An international Code of Practice for Dosimetry Based on Standards of Absorbed dose to Water // IAEA. Technical Report Series. – № 398. – Vienna, 2000.

9. Bocanek, J. Report from the feasibility testing of MatriXX device for RapidArc QA / J. Bocanek // Proceed of Varian Medical Systems workshop. – 2008.

10. Van Esch, A. Dominique P. Huyskens Mauro Iori Christian Clermont and Magali Devillers On-line quality assurance of rotational radiotherapy treatment delivery by means of a 2D ion chamber array and the Octavius phantom / A. Van Esch, D.P. Huyskens, M. Iori // Med. Phys. – V. 34. – 2007. – P. 3825–3837.

11. Depuydt, T. A quantitative evaluation of IMRT dose distributions: refinement and clinical a ssessment of the gamma evaluation / T. Depuydt A. Van Esch, D.P. Huyskens // Radiotherapy and Oncology. – V. 62. – 2002. – P. 309–319.

12. Van Esch, A. Acceptance tests and quality control (QC) procedures for the clinical i mplementation of intensity modulated radiothe rapy (IMRT) using inverse planning and the sliding window technique: experience from five radiotherapy departments / A. Van Esch, J. Bohsungb, P. Sorvaric [et al.] // Radiotherapy and Oncology. – V. 65. – 2002. – P. 53–70.


Review

For citations:


Tsitovich Y.V., Hmyrak A.I., Tarutin A.I., Kiselev M.G. METHODS AND HARDWARE OF DOSE OUTPUT VERIFICATION FOR DYNAMIC RADIOTHERAPY. Devices and Methods of Measurements. 2013;(1):97-102. (In Russ.)

Views: 901


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2220-9506 (Print)
ISSN 2414-0473 (Online)