Preview

Evaluation of the Magnet Breakaway Force Measurement Accuracy of the NT-800 Sensors for Early Detection of Defects of Their Manufacturing

https://doi.org/10.21122/2220-9506-2021-12-3-230-238

Abstract

Сontrol of mechanical stresses formed with the deposition of nickel coatings plays an important role in the diagnosis of coatings’ technical condition. Large internal stresses can lead to cracking or flaking of coatings which is completely unacceptable for critical parts and assembly units used, for example, in space technology for which reliability is of paramount importance. An important aspect of internal stresses monitoring is the measurement error of the instruments used. The purpose of this work was to determine the characteristics of the device sensors, which make the assessment of their manufacturing possible at the preliminary stage of the measuring equipment assembling in order to maintain the required accuracy of subsequent measurements.

In most cases the measurement error assessment is possible only after the equipment manufacture and calibration. In this paper it is proposed to evaluate the accuracy characteristics of device sensors based on the precision (repeatability and reproducibility) of the primary informative parameter recording. In the case of the NT-800 device that was developed at the Institute of Applied Physics of the National Academy of Sciences of Belarus the effect of precision characteristics deterioration on the eventual measurement error is demonstrated. Determining the precision parameters before establishing correlation dependences between the primary informative parameter and the measured characteristic is proposed in order to reject poorly manufactured sensors and reduce labor costs.

In particular, measurements of the magnitude proportional to the magnetic breakaway force were carried out using the NT-800 device with nickel specimens simulating coatings with a thickness of 200 to 700 μm and a rolling value from 0 to 40 %. It was established that in the case of well-made sensors the variation coefficient calculated from the dispersion of repeatability is in the range 0.2–0.6 %, and the variation coefficient calculated from the dispersion of reproducibility does not exceed 0.9 %. In the case of a sensor with the sensitive element parameters worsened, the variation coefficient of repeatability and reproducibility were up by one and a half times. Deterioration of the precision characteristics resulted in significant changes in the readings of the calibrated instrument. Thus the absolute measurement error for a sensor with a poorly made sensitive element turned out to be approximately 3 times higher in the range of 200– 300 MPa than that for a sensor with good precision parameters.

About the Authors

A. Yu. Kutsepau
Institute of Applied Physics of the National Academy of Science of Belarus
Belarus

Akademicheskaya str., 16, Minsk 220072



A. P. Kren
Institute of Applied Physics of the National Academy of Science of Belarus
Belarus

Address for correspondence: Kren A.P. Institute of Applied Physics of the National Academy of Science of Belarus, Akademicheskaya str., 16, Minsk 220072, Belarus
e-mail:7623300@gmail.com



Y. V. Hnutsenka
Institute of Applied Physics of the National Academy of Science of Belarus
Belarus

Akademicheskaya str., 16, Minsk 220072



References

1. Practical residual stress measurement methods / edited by Gary S. Schajer: John Wiley & Sons Ltd, 2013, 310 p.

2. Rasilo P., Aydin U., Martin F., Belahcen A., Kouhia R., Daniel L. Equivalent strain and stress models for the effect of mechanical loading on the permeability of ferromagnetic materials. IEEE Transactions on Magnetics, 2019, vol. 55, iss. 6, pp. 1‒4. DOI: 10.1109/TMAG.2018.2890407

3. Weisensel N., Liang C. Magnetotagging, A New method for active and passive nondestructive evaluation of stresses and defects in composite structures. Materials Technology, 2016, vol. 10, iss. 7‒8, pp. 142‒144. DOI: 10.1080/10667857.1995.11752613

4. Ivanova Y., Partalin T. Comparative measurements of the stress state in a rolled carbon steel using magnetic Barkhausen noise and ultrasonic method. Russian Journal of Nondestructive Testing, 2012, no. 48, pp. 137–146. DOI: 10.1134/S1061830912020040

5. Filinov V.V., Kuznetsov A.N., Arakelov P.G. Monitoring stressed state of pipelines by magnetic parameters of metal. Russian Journal of Nondestructive Testing, 2017, no. 53, pp. 51–61. DOI: 10.1134/S1061830917010065

6. Rudnitskiy V.A., Hnutsenka E.V., Kren A.P. Opredelenie tolshchiny dvuhslojnyh nikel'-hromovyh pokrytij magnitnym metodom [Determination of the thickness of double-layered nickel-chrome platings using magnetic method]. Metrologiya i priborostroyenie [Metrology and Instrumentation], 2015, no. 4, pp. 24–29 (in Russian).

7. Gorkunov E.S., Povolotskaya A.M., Zadvorkin S.M., Putilova E.A. Comparative analysis of the magnetic characteristics of plastically deformed metal in different zones of a welded pipe under elastic deformation. Russian Journal of Nondestructive Testing, 2017, vol. 53, pp. 636–643. DOI: 10.1134/S1061830917090066

8. Wilrich P.-T. Critical values of Mandel’s h and k, the Grubbs and the Cochran test statistic. Advances in Statistical Analysis, 2013, no. 97(1), pp. 1‒10. DOI: 10.1007/s10182-011-0185-y


Review

For citations:


Kutsepau A.Yu., Kren A.P., Hnutsenka Y.V. Evaluation of the Magnet Breakaway Force Measurement Accuracy of the NT-800 Sensors for Early Detection of Defects of Their Manufacturing. Devices and Methods of Measurements. 2021;12(3):230-238. https://doi.org/10.21122/2220-9506-2021-12-3-230-238