APPLICATION OF THE SPECTROMETRIC METHOD FOR CALCULATING THE DOSE RATE FOR CREATING CALIBRATION HIGHLY SENSITIVE INSTRUMENTS BASED ON SCINTILLATION DETECTION UNITS

https://doi.org/10.21122/2220-9506-2017-8-3-246-253

Abstract

Devices based on scintillation detector are highly sensitive to photon radiation and are widely used to measure the environment dose rate. Modernization of the measuring path to minimize the error in measuring the response of the detector to gamma radiation has already reached its technological ceiling and does not give the proper effect. More promising for this purpose are new methods of processing the obtained spectrometric information. The purpose of this work is the development of highly sensitive instruments based on scintillation detection units using a spectrometric method for calculating dose rate.

In this paper we consider the spectrometric method of dosimetry of gamma radiation based on the transformation of the measured instrumental spectrum. Using predetermined or measured functions of the detector response to the action of gamma radiation of a given energy and flux density, a certain function of the energy G(E) is determined. Using this function as the core of the integral transformation from the field to dose characteristic, it is possible to obtain the dose value directly from the current instrumentation spectrum. Applying the function G(E) to the energy distribution of the fluence of photon radiation in the environment, the total dose rate can be determined without information on the distribution of radioisotopes in the environment.

To determine G(E) by Monte-Carlo method instrumental response function of the scintillator detector to monoenergetic photon radiation sources as well as other characteristics are calculated. Then the whole full-scale energy range is divided into energy ranges for which the function G(E) is calculated using a linear interpolation.

Spectrometric method for dose calculation using the function G(E), which allows the use of scintillation detection units for a wide range of dosimetry applications is considered in the article. As well as describes the method of calculating this function by using Monte-Carlo methods and the features of its application. The results of the calculation function G(E) for the detection unit on the basis of NaI(Tl) detector (Ø40 mm, h = 40 mm) to use it as a comparator for kerma rate in the air certification of low intenseе photon radiation fields. 

About the Authors

R. V. Lukashevich
ATOMTEX
Belarus

Address for correspondence: Lukashevich R. –  SPE «ATOMTEX», Gikalo str., 5, Minsk 220005, Belarus   e-mail: rimlianin@gmail.com



G. A. Fokov
ATOMTEX
Belarus


References

1. Moriuchi S. A new method of dose evaluation by spectrum-dose conversion operator and determination of the operator. JAERI 1209, Japan Atomic Energy Research Institute, 1970 (in Japanese).

2. Moriuchi S., Nagoya T., Sakamoto S., Saito K. Determination of spectrum-dose conversion operator for spherical NaI(Tl) scintillators, JAERI-M 8092, Japan Atomic Energy Research Institute, 1972 (in Japanese).

3. Moriuchi S., Satsuma M., Saito K. Development of a dissymmetric system using spectrometric technique suitable for operational radiation dose measurements and evaluation. Proc. 10-th IRPA International Congress, Hiroshima, May 2000, p-3b-197.

4. Fominykh V.I., Shulgovich G.I., Kozhemiakin V.A. Measurement of small levels of gamma radiation by spectrometric method using the «spectrum-dose» operator. Metrologija [Metrology], no. 10, 1983, pp. 32–39. (in Russian)

5. Aoyama K., Masui K., Yamamura S., Nakamura T., Yabutani T., Namito Y. Development of low-energy x- ray survey-meter, radiation detectors and their uses. KEK Proceedings, 2006, 7 November, pp. 57–67.

6. Yamamura Seini, Nakamura Takashi, Itou Katsuhito, Hatakeyama Osamu, Masui Kaoru. Development of wide-energy range X/gamma-ray survey-meter. Journal of Nuclear Science and Technology, no. 45, iss. 5, 2008, pp. 187–190. doi: 10.1080/00223131.2008.10875818

7. Babenko V.V., Isaev A.G. Spectrometric method for the measurement of the absorbed dose of gamma radiation using a NaI(Tl) detector. Spektrometricheskiy analiz. Apparatura i obrabotka dannykh na PEVM [Proc. XI annual seminar «Spectrometric Analysis. Equipment and data processing on PC»]. Obninsk, Federal State Educational Institution «GICPK» Publ., 2005, pp. 67–77 (in Russian).

8. Cho G., Kim H.K., Woo H. Electronic dose conversion technique using a NaI(Tl) detector for assessment of exposure dose rate from environmental radiation. IEEE Transactions on Nuclear Science, no. 45, 1998, pp. 981– 985. doi:10.1109/23.682692

9. Moriuchi S., Tsutsumi M., Saito K. Construction of Response Matrices for Various Cylindrical and Spherical NaI(Tl) Scintillation Detectors for Gamma Rays and the Test Results. Japanese Journal of Health Physics, no. 44, 2009, pp. 122–133.

10. Fokov G.A., Shulgovich G.I. Calculation of the instrumental response functions of a standard NaI gamma-ray detector using the universal program code SNEGMONT. Spektrometricheskiy analiz. Apparatura i obrabotka dannykh na PEVM [Proc. XI annual seminar «Spectrometric Analysis. Equipment and data processing on PC»]. Obninsk, Federal State Educational Institution «GICPK» Publ., 2008, pp. 145–158 (in Russian).

11. Dozimetricheskaya ustanovka gamma-izlucheniya UDG-AT110 [АТ110 Gamma Beam Irradiator with Calibration Bench]. Available at:: http://www.atomtex.com/sites/default/files/udg-at110.pdf (accessed 04.07.2017)


Review

For citations:


Lukashevich R.V., Fokov G.A. APPLICATION OF THE SPECTROMETRIC METHOD FOR CALCULATING THE DOSE RATE FOR CREATING CALIBRATION HIGHLY SENSITIVE INSTRUMENTS BASED ON SCINTILLATION DETECTION UNITS. Devices and Methods of Measurements. 2017;8(3):246-253. (In Russ.) https://doi.org/10.21122/2220-9506-2017-8-3-246-253