Preview

INCREASING OF ACCURACY FOR DENSITY MEASUREMENT MAKING BY A RADIOISOTOPIC BASIS WEIGHT SENSOR

Abstract

The experimental investigations of spatial energy distribution for beta particles in a measuring gap of a basis weight sensor are described. We propose the measurement procedure, which employs a local domain of the ionization chamber input window. The average scattering angles in polyethylenterephtalate films are determined. Suggested technical solutions allow more then three times to reduce the error of measuring of radioisotope densitometer.

For citations:


Ermakovich O.L., Lisovsky G.A., Kuchynski P.V., Titovitsky I.A. INCREASING OF ACCURACY FOR DENSITY MEASUREMENT MAKING BY A RADIOISOTOPIC BASIS WEIGHT SENSOR. Devices and Methods of Measurements. 2015;6(1):70-75. (In Russ.)

Views: 722


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 2220-9506 (Print)
ISSN 2414-0473 (Online)